Abstract

Small sample 14C dating is tested using conventional as well as high-resolution low-level liquid scintillation (LS) spectrometers. Contrasted are the results obtained dating ∼25, 125 and 250mg of elemental carbon in standard size counting vials (3mL) and 0.3mL teflon and quartz vials. It is demonstrated that the improved performance of the 0.3mL counting vials enables just adequate resolution of 25mg and very good resolution of 100mg carbon samples both at Modern and Old age limits when the determination is made in a high-resolution low-level LS spectrometer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.