Abstract

130 Male Wistar rats, 2-3 months old and weighing 250-350 g were operated on to implant biodegradable small-calibre vascular prostheses (length 10 mm; internal diameter 1.5 mm) in their infrarenal abdominal aorta. The mean operation time was 40 min, the mean aortic cross-clamping time 25 min. The early patency rate was 100%, the late patency rate was 97.7%, and the operative mortality was 3.1%. Microscopical examination of the biodegradable prostheses from 1 h up to 1 year after implantation demonstrated reproducible morphological results; in these prostheses a new arterial wall regenerated which had a structure very similar to the normal arterial wall. It was concluded that the rat is an appropriate experimental laboratory animal for testing new types of small-calibre vascular prostheses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.