Abstract

AbstractThe purpose of this study is to present a novel small animal PET imaging system, which can be integrated with a small animal CT or MRI imaging in a PET/CT or PET/MRI schemes. The specific small animal PET imaging design consists of two planar detectors, with LYSO crystals, followed by Position Sensitive Photomultiplier tubes (PSPMTs). The output of the each PSPMTS is led to the corresponding modern electronics for analysis and digitization while at the end is stored on a host computer for further processing. In order to evaluate the average performance of the aforementioned PET system, SIMSET simulation tool was used. So, the small animal PET imaging system was simulated, taking into account that the radioactive source inside it’s field of view was a Derenzo-like phantom full of FDG. Attenuation and scatter phenomena, as well as random events were also considered during simulation process. The simulated data (Derenzo-like phantom sinogram) was then reconstructed with the ordered subsets versions of EM-ML, ISRA and WLS algorithms, considering regularization and penalized schemes. Regularization techniques accelerate reconstruction procedure while penalized methods increase signal to noise ratios with small image resolution degradation. Reconstructed images are further analyzed and evaluated so that a general conclusion on final system resolution output to be obtained.KeywordsSmall animal imagingDedicated small animal PET systemsIterative image reconstructionContrast to noise ratio