Abstract

The reduced background of 2 nSv h(-1) of the Slănic-Prahova Low Background Radiation Laboratory allowed installing a calibration stand for low-dose-rates dosimetry. The stand is provided with (60)Co, (137)Cs and (241)Am low activity sources. A Physikalisch-Technische Bundesanstalt-calibrated AUTOMESS 6150 AD-6 dose rate meter with a 6150 AD-b/H external probe was used to check to what extent this stand could serve as a low background calibration facility. A detailed analysis of possible uncertainties in measuring dose rates evidenced an extended uncertainty related to the certified calibration as well as instrument readings of about 3 % for a confidence level of 95 %. In these conditions, the experimentally determined dose rates for all three gamma ray sources and for source-to-probe distances varying between 0.3 and 2.0 m confirmed a good correlation between the calculated and measured dose rates. At the same time, dose rates perfectly obey to an inverse squared distances law.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.