Abstract

To assess the influence of high temporal resolution on the perfusion measurements and image quality of perfusion maps, by applying simultaneous-multi-slice acquisition (SMS) dynamic susceptibility contrast-enhanced (DSC) magnetic resonance imaging (MRI). DSC-MRI data using SMS gradient-echo echo planar imaging sequences in 10 subjects with no intracranial abnormalities were retrospectively analyzed. Three additional data sets with temporal resolution of 1.0, 1.5, and 2.0s were created from the raw data sets of 0.5s. Cerebral blood flow (CBF), cerebral blood volume, mean transit time (MTT), time to peak (TTP), and time to maximum tissue residue function (T max) measurements were performed, as was visual perfusion map analysis. The perfusion parameter for temporal resolution of 0.5s (reference) was compared with each synthesized perfusion parameter. CBF, MTT, and TTP values at temporal resolutions of 1.5 and 2.0s differed significantly from the reference. The image quality of MTT, TTP, and T max maps deteriorated with decreasing temporal resolution. The temporal resolution of DSC-MRI influences perfusion parameters and SMS DSC-MRI provides better image quality for MTT, TTP, and T max maps.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.