Abstract

Currently, robot-assisted surgical systems are used to reduce the error range of total knee arthroplasty (TKA) osteotomy and component positioning. We used 20 sawbone models of the femur and 20 sawbone models of the tibia and fibula to evaluate the osteotomy effect of 'Skywalker' robot-assisted TKA. The maximal movement of the cutting jig was less than 0.25mm at each osteotomy plane. The mean and standard deviation values of the angle deviation between the planned osteotomy plane and the actual osteotomy plane at each osteotomy plane were not more than 1.03° and 0.55°, respectively. The mean and standard deviation values of absolute error of resection thickness at each osteotomy position were less than 0.78 and 0.71mm, respectively. The 'Skywalker' system has good osteotomy accuracy, can achieve the planned osteotomy well and is expected to assist surgeons in performing accurate TKA in clinical applications in future.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.