Abstract

Photoacoustic (PA) signal is highly dependent on the structure of the target. In the present study, PA imaging system with parabolic array transducer for multi-angle detection is developed and the skin vasculature is observed. The transducer was consisted of 256 ch 1-3-composite elements with the center frequency of 10 MHz. The diameter was 42.4 mm and the opening angle was 90 degree. A hole was made in the center of the transducer for transmission of the laser. PA signal was acquired by a programmable ultrasound data acquisition system with the sampling frequency of 60 MHz. A diode laser with the wavelength of 532 nm, pulse width of 8 ns, pulse energy of 5 mJ, and repetition frequency of 10 Hz was equipped for generation of PA signal. Three dimensional image was reconstructed by delay-and-sum beamforming of PA signals received by all channels. The frame rate for displaying x-y, y-z, and x-z images was 10 Hz, and the full width at half maximum (FWHM) of the x-z image was 70 μm. 3D structure of the skin ca...