Abstract

To determine how the skin dose varies in patients receiving radiation treatment for breast cancer in the prone and supine positions. Fifty patients were scanned in the prone and supine positions. A radiation treatment plan was created for the left breast using a 6-MV beam for a prescribed dose of 42.66Gy in 16 fractions. The dose was calculated using 1- and 2.5-mm calculation grid sizes and the surface dose was compared in both techniques. The median gantry angles relative to the skin surface at the central axis were 8 and 52 degrees for treatment in the prone and supine positions, respectively. The mean dose difference between the prone and supine techniques was statistically significant from 3- to 5-mm depth for both grid sizes. For the 1-mm calculation grid size, the doses at 3-, 4-, and 5-mm depths in the prone and supine techniques were 87.80% and 89.10% (P<0.003), 91.92% and 94.50% (P<0.00), and 95.30% and 98.20% (P<0.00), respectively; for the 2.5-mm grid size, the respective doses were 87.10% and 88.59% (P<0.00), 91.60% and 94.63% (P<0.00), and 95.10% and 97.80% (P<0.00), respectively. This study demonstrates that the prone technique facilitates a relatively lower skin dose than the supine technique. This observation is probably due to the beam angle. The beam is more perpendicular to the skin surface in the prone technique, whereas it is more tangential in the supine technique, which may deliver a higher skin dose. Thus, the dose to the skin should be evaluated in the prone technique, and if desired, the skin dose could be carefully augmented via a bolus or beam spoiler.