Abstract
Abstract Superficial dose from 6- and 18-MV photon beams has been studied by measuring surface dose and shallow build-up dose using radiographic film EDR2, radiochromic film EBT2 and plane-parallel chamber. Measurements have been made for intensity- and non-intensity-modulated beams. The results show that the surface dose was found to be 19.8% and 10% of maximum dose in unmodulated fields for 6 and 18 MV photon beams, respectively. The study further showed that intensity modulation decreased surface dose by 1.1% and 0.7% for the same field size at 6 and 18 MV, respectively, and surface dose was dropped by magnetically sweeping contaminating electrons. EDR2 and EBT2 films show in good agreement in shallow build-up region. This study demonstrated the capability of EDR2 film, in addition to radiochromic film, to measure surface and build-up dose in case of treatment planning system uncertainties with regard to skin toxicity or shallow target coverage.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.