Abstract
The 2.5 MV Imaging beam produced by a Varian TrueBeam linear accelerator produces a dose build up effect at the beam entrance similar to other high energy photon beams. The surface dose values were found to range from 39% of maximum dose at a 5cm × 5cm field size up to 69% of maximum at a 40cm × 40cm field. The depth of maximum dose deposition was found to range from 5mm at smaller field sizes to 4mm at larger field sizes. Whilst large absorbed doses will not be delivered utilizing these beams, the data provided will allow the medical physics community to assess and estimate doses to patient's skin and subcutaneous tissue from low energy MV imaging beams.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Australasian physical & engineering sciences in medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.