Several techniques for determining the size of small fluid microdrops with diameters ranging from 5 to 30 μm have been developed and evaluated using an automated variation on the Millikan oil drop experiment. The average diameter of a large sample of monodisperse fluid drops was determined by measuring their terminal velocity in air, or if charged, their motion under the influence of an electric field, as well as by measurement of the magnitude of their Brownian motion. The diameter of individual drops was determined optically, by direct observation using an imaging system based on a charge-coupled-device camera. The technique used to analyze the image data is based on a best-fit technique taking the point spread function of the lens into account, and yields results accurate to 1% (based on a single image) without the need for any calibration. By combining this technique with terminal velocity measurements, the density of the fluid can be determined to similar accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.