Abstract

A laboratory micro X-ray fluorescence spectrometer based on a special polycapillary X-ray lens (PXRL) was used to carry out the source apportionment of aerosol particles. In the curve of the distribution of the X-ray intensity in the focal spot of the special PXRL, there was a plateau with a diameter of 21.3 μm in which the distribution of the X-ray intensity was homogeneous. The gain in flux density in the plateau of the PXRL is 1490. The uniformity of this plateau was 2.9%. This was helpful for the quantitative X-ray fluorescence analysis of a single aerosol particle with smaller size than that of the plateau of the PXRL. The fingerprint database of aerosol particles with given sizes from various air pollution sources was established with the single particle analysis method. The size-resolved source apportionment of aerosol particles in haze in Beijing city was performed with this fingerprint database.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.