Abstract

We describe a newly constructed annular-type differential mobility analyzer (DMA) with an effective electrode length of 60 cm, which is longer than that of our original DMA (40 cm length). This long DMA was developed to extend the classification size of particles measuring up to 1.5 μm. As an application of this DMA, the mean diameters and standard deviations were determined for six samples of monodisperse polystyrene latex (PSL) particles ranging from 0.94 to 1.27 μm in nominal diameter. These PSL particles suspended in double-distilled water were aerosolized by a glass nebulizer and then introduced into the DMA. The mean diameters and standard deviations of these PSL particles were calculated by Ehara's method and compared with the nominal diameters and uncertainties. There was good correlation between the nominal diameters of these samples—particularly for recently certified samples—and the measured diameters. Classification of aerosol particles more than 1 micrometer in diameter using this DMA will be useful for many purposes.