Abstract

In the present work, the question how to obtain high-quality laser diffraction (LD) results is discussed by investigating various hurdles that can be encountered during particle size measurements in wet dispersions and the associated data interpretation. Following this an effective troubleshooting is discussed based on theoretical insight into the LD measurement. As an important element in the Mie theory, the refractive indices of the model compounds, bedaquiline and cinnarizine, were quantified using the LD software, the Becke line technique, and the single solvent technique, as described by Saveyn et al. [Saveyn H, et al. Part Part Syst Char. 2001;19:426-432]. The influence of parameters such as obscuration level, background quality, and fitting of data were investigated and a summative flow chart has been provided. Through this analysis the present work emphasizes the need for a systematic approach when conducting LD measurements, including standard performance of an obscuration titration and an extended method optimization, to reach high-quality LD results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.