Abstract

We propose an integrating sphere photometer with six detection ports for total luminous flux measurement, which significantly improves the uniformity of spatial response compared to the conventional single-port detection design. Numerical simulations based on the geometric radiative transfer equation show that a spatial response distribution function of the new design is uniform within 2% with respect to all spatial directions. The related spatial mismatch error is calculated to be less than 0.3% for all the realistic cases of angular intensity distribution of a test lamp. As a result, the new design practically eliminates the spatial mismatch error of an integrating sphere photometer, so that a high-accuracy measurement can be achieved without the complicated spatial mismatch correction procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.