Abstract
The surgical repair of type A aortic dissection often involves prosthetic proximal aortic and arch reconstruction. Hypothermic circulatory arrest is typically used in these complex surgeries given the required prolonged ischaemia and the associated morbidity and mortality. A novel vascular anastomoses device (Device) has been developed to rapidly connect a native vessel to a polyester graft. This study describes deployment of the Device in the ovine model (n = 3; 6 carotid arteries). Anastamoses were created rapidly, and brain ischaemia time was limited to 6 min in all but one vessel. All vessels remained fully patent with normal blood flow and thrombus-free transitions through 6 months. Results thus suggest that this Device has the potential to reduce anastomosis time versus conventional suturing techniques and thereby reduce hypothermic circulatory arrest time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.