Abstract

The need to adapt optimal conditions of sirolimus blood level monitoring in laboratories led us to optimize an high-performance liquid chromatography-ultraviolet method and compare the elution performances using the mobile phase A, 68% MeOH/2% acetonitrile (ACN)/30% H(2)O and mobile phase B, 30% MeOH/42% ACN/28% H(2)O. Samples were assayed with 1-chlorobutane, redissolved in MeOH/water and injected onto a C-18 column at 50 degrees C. The assay achieved sensitivity of 2.5-150 ng/ml (CV = 10.6%) and recovery of 92-103.6%. The intra- and interassay precisions ranged from 3.3% to 13% and from 5.9% to 15% for quality controls of 7.5, 60 and 120 ng/ml. The mobile phase A was unable to elute and recover sirolimus and internal standard in the expected retention time and concentration. Under our working conditions, the assay was precise, accurate and sensible, stressing the importance of establishing for the best working conditions according to the staff and demands of the laboratory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.