Abstract

Cell-free DNA (cfDNA) content in plasma has been studied as a biomarker for sepsis. Recent publications show that the cfDNA content in sepsis patients entering intensive care unit who were likely to survive had a total cfDNA concentration of 1.16 ± 0.13μg/mL compared to 4.65 ± 0.48μg/mL of non-survivors. Current methods for measuring cfDNA content in plasma were designed to amplify and measure low concentrations of specific DNA, making them unsuitable for low-cost measurement of total cfDNA content in plasma. Here, we have developed a point of care (POC) device that uses a thread silicone device as a medium to store a fluorescent dye which eliminates the need for preparatory steps, external aliquoting and dispensing of reagents, preconcentration, and external mixing while reducing the detection cost. The device was paired with a portable imaging system with an excitation filter at 472 ± 10nm and an emission filter of 520 ± 10nm that can be operated with just 100mA current supply. The device was demonstrated for use in the quantification of buffered cfDNA samples in a range 1-6μg/mL with a sensitivity of 5.72 AU/μg/mL and with cfDNA spiked in plasma with a range of 1-3μg/mL and a sensitivity of 5.43 AU/μg/mL. The results showed that the device could be used as a low-cost, rapid, and portable POC device for differentiating between survivors and non-survivors of sepsis within 20min.