Abstract

A new reagent for the chromogenic Limulus amebocyte lysate (LAL) assay is described. LAL was formulated for optimal performance in either an endpoint procedure or a kinetic procedure with the chromogenic substrate, buffer, and LAL components colyophilized as a single reagent. The kinetic chromogenic method required an incubating microplate reader coupled to a computer for collection and analysis of data. The kinetic method had a longer incubation time than the endpoint method and spanned a range of over 3 orders of magnitude compared with the 1-order-of-magnitude range of the endpoint assay. The kinetic method was less subject to operator error, since readings were continuous and automatic. The endpoint test was more operator intensive, requiring both addition of acetic acid to stop the reaction and transfer of the sample to the reading device. A single-step chromogenic reagent was also prepared without lyophilization by mixing reconstituted gel clot LAL with a buffer and a chromogenic substrate. The reagent prepared in this manner performed as well as the colyophilized agent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.