Abstract

A new phase-encoding scheme for gradient- and spin-echo (GRASE) imaging giving a short effective TE is described. Unlike previous orders, phase encoding is centric rather than sequential. The sequence is a development of k-banded GRASE that uses different time segments of the echo train to encode different bands of k space. This phase-encoding order has been implemented in single-shot sequences on an imager with high performance gradients. Approximately 144 phase-encoding lines can be acquired in an echo train time of 390 ms. With centric phase encoding, the effective TE is 8 ms, compared with 75 ms for sequential encoding, and signal-to-noise ratios (SNRs) in brain tissue are 50 to 70% higher. The sequence can be employed in, for example, diffusion and velocity imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.