Abstract

Traditionally, Q measurement requires a nonzero frequency bandwidth or time period. Contrasted to this, the principle of a new single-frequency relative Q measurement method is developed. It is found that Q is directly or inversely proportional to the normalized input resonant resistance if a moderate perturbation condition is satisfied. Theoretical proof and experimental verification of the single-frequency method's validity are presented. Consequently, a relative and, often used in dielectric measurements, can be measured using a much simpler measurement system. Moreover, error analysis shows that, in making a relative Q measurement, the error in the single-frequency method is smaller than that in the traditional bandpass method when using a reflectometer.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.