Abstract
Endocardial (ENDO) or epicardial (EPI) pacing systems are implanted in infants but it remains unclear which system should be preferred. We evaluated the results of children <or=1 year who underwent pacemaker (PM) implantation at our centre with a retrospective analysis. Between 1992 and 2004, 56 patients, 37 of whom had other congenital heart defects (CHDs), received a PM at 4.4 +/- 3.8 months of age for atrioventricular block (n = 52) and sinus node dysfunction. Rate-responsive ventricular demand pacing (VVIR) PMs were implanted in 25 patients (19 ENDO), dual-chamber demand pacing (DDD) in 29, and rate-responsive atrial demand pacing (AAIR) in 2 (all EPI). Follow-up (FU) was 4.5 +/- 3.5 (range 0.3-13) years: 15 pacing system failures occurred among the 56 patients (26%) after 4.5 +/- 3.2 years, with a significantly reduced success rate for EPI (21-fold increase of the risk of failure) and complex CHD. Also in patients without surgery for CHD, EPI showed a worse outcome. Among the 91 leads implanted, failures occurred more significantly in EPI (18% of atrial, 24% of ventricular leads) than in ENDO (5% of ventricular leads). No venous occlusion was found at FU. Single-lead, VVIR ENDO pacing had higher efficiency and safety than EPI, and it might be the best choice for PM implantation in infants. However, because of small patient numbers and lack of longer FU, these findings should be treated with caution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.