Abstract

A single-tone mid-infrared frequency modulation (MIR-FM) spectrometer consisting of a cw-OPO-based laser system, a 500 MHz LiTaO 3 electro-optical modulator (EOM), and a high-bandwidth GaAs mid-infrared detector has been developed. In order to assess the instrument's sensitivity and time resolution, FM spectra of selected CH 4 transitions around 3070 cm -1 were measured and the reaction Cl + CH 4 following the 193 nm excimer laser photolysis of oxalyl chloride was investigated by recording concentration-time profiles of HCl at 2925.90 cm -1 in a low-pressure slow-flow reactor. Furthermore, OH radicals were generated by UV photolysis of H 2O 2 and its transients were recorded at 3447.27 cm -1. The minimal detectable absorption of the spectrometer was determined to be A min=4⋅10-4 (Δ f BW=1 MHz, ν~=3447 cm -1) by using the Allan approach. Mainly due to thermal noise contributions of the easy-to-saturate photodetector, the detection limit is about a factor of 4 above the shot-noise limit. To the best of our knowledge, this work reports the first implementation of a single-tone MIR-FM spectrometer based on an external EOM modulation scheme and its use for the detection of transient molecular species.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call