Abstract
Talbot–Lau x-ray imaging provides additionally to the conventional attenuation image, two further images: the differential phase-contrast image which is especially sensitive to differences in refractive properties and the dark-field image which is showing the x-ray scattering properties of the object. Thus, in the dark-field image sub-pixeled object information can be observed. As it has been shown in recent studies, this is of special interest for lung imaging. Changes in the alveoli structure, which are in the size of one detector pixel, can be seen in the dark-field images. A fast acquisition process is crucial to avoid motion artifacts due to heartbeat and breathing of the patient. Using moiré imaging the images can be acquired with a single-shot exposure. Nevertheless, the spatial resolution is reduced compared to the phase-stepping acquisition. We evaluate the results of both imaging techniques towards their feasibility in clinical routine. Furthermore, we analyse the influence of artificial linear object movement on the image quality, in order to simulate the heartbeat of a patient.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.