Abstract

PurposeTo analyze single session, intrauser reliability of a Scheimpflug device for anterior chamber (AC) and corneal parameters. MethodsIn this observational study, 100 normal candidates underwent Scheimpflug analysis with Sirius 3D Rotating Scheimpflug Camera and Topography System (Costruzione Strumenti Oftalmici, Italy). Two scans in dark room conditions were performed by the same experienced user. The candidates were asked to keep both eyes closed for 5min before the scans. Exclusion criteria were previous ocular surgery, corneal scarring and anterior segment/posterior segment anomalies. Only the right eyes were used for the analysis. Both corneal (central, minimum, and apical thickness, volume, horizontal visible iris diameter, and apical curvature) and anterior chamber (volume, depth, angle, horizontal diameter) measurements were evaluated. ResultsThere was no difference in the means of repeated measurements (p>0.05, ANOVA). Intraclass correlations between the measures were high and ranged from 0.995–0.997 for corneal to 0.964–0.997 for anterior chamber (AC) parameters. The precision of repeatability measures (1.96×Sw) was approximately 5μ for the central and minimum corneal thickness, 8μ for the apical corneal thickness, 0.06mm for AC (anterior chamber) depth and less than 2° for the AC angle. ConclusionsSirius Scheimpflug system has high repeatability for both corneal and AC parameters in normal eyes.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call