Abstract

Laparoendoscopic single-site surgery (LESS) allows better cosmesis and decreased pain when compared with traditional laparoscopy (TL). Instrument crowding and diminished triangulation are limitations. This study evaluates a novel single-port surgical platform (NSP) designed to facilitate the transition from TL to LESS (TransEnterix SPIDER, Durham, NC). NSP and TL were compared using standardized dry lab tasks. Feasibility of NSP in gynecology was assessed using the porcine model. Completion times with NSP were longer for ring transfer (P = .025) and trended longer for disc cutting (P = .074). Disc cutting accuracy was lower with NSP versus TL (P = .008). NSP operative times for hysterectomy, pelvic and para-aortic node dissection were 22.3, 13.3, and 26 minutes, respectively, without complications. Fatigue and lack of integrated bipolar cautery were limitations of NSP. While application of NSP to gynecology appears feasible, lower performance when compared with TL underscores the need for caution when applying NSP to humans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.