Abstract
The application of soft ionization methods for mass spectrometry (MS), such as single-photon ionization (SPI) using vacuum ultraviolet (VUV) light, provides powerful analytical instrumentation for real-time on-line monitoring of organic substances in gaseous matrixes. A compact and mobile quadrupole mass spectrometer (QMS) system using a novel electron beam pumped rare gas VUV lamp for SPI has been developed for on-line analysis of organic trace compounds (ppb concentrations). The VUV radiation of the light source is employed for SPI in the ion source of the QMS. The concept of the interfacing of the VUV light source with the QMS is described and the SPI-QMS is characterized. On-line detection limits down to 50 ppb for benzene, toluene, and m-xylene were achieved. The instrument is well suited for continuous measurements of aromatic and aliphatic trace compounds and can therefore be used for on-line monitoring of trace compounds in dynamically fluctuating process gases. First measurements of gas standards, petrochemical samples, and on-line monitoring of automotive exhaust are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.