Abstract
Digital polymerase chain reaction (PCR) is a promising technique for estimating target DNA copy number. PCR solution is distributed throughout numerous partitions, and following amplification, target DNA copy number is estimated based on the proportion of partitions containing amplified DNA. Here, we identify approaches for obtaining reliable digital PCR data. Single molecule amplification efficiency was significantly improved following fragmentation of total DNA and bias in copy number estimates reduced by analysis of short intact target DNA fragments. Random and independent distribution of target DNA molecules throughout partitions, which is critical to accurate digital PCR measurement, was demonstrated by spatial distribution analysis. The estimated relative uncertainty for target DNA concentration was under 6% when analyzing five digital panels comprising 765 partitions each, provided the panels contained an average of 212 to 3,365 template molecules. Partition volume was a major component of this uncertainty estimate. These findings can be applied to other digital PCR studies to improve confidence in such measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.