Abstract

The study of acoustic scattering by single microbubbles has the potential to offer improved signal processing techniques. A microacoustic system that employs a hydrodynamically-focused flow was used to detect radiofrequency (RF) backscatter from single microbubbles. RF data were collected using a commercial scanner. Results are presented for two agents, namely Definity ® (Lantheus Medical Imaging, N. Billerica, MA, USA) and biSphere ® (Point Biomedical Corp, San Carlos, CA, USA). The agents were insonified with amplitude-modulated pulses, and it was observed in both agents that a subpopulation of microbubbles did not produce a measurable echo from the first-half amplitude pulse, but did produce a response from the full amplitude pulse and from a subsequent half amplitude pulse. The number of microbubbles in this subpopulation was seen to increase with increasing transmit amplitude. These results do not bear out the simple theory of microbubble–pulse sequence interaction and invite a reassessment of signal processing approaches. (E-mail: d.h.thomas@ed.ac.uk)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.