Abstract
Laser-induced fluorescence is used as a diagnostic tool for testing the tuning between a longitudinal mode of a nitrogen-pumped dye laser and a uranium spectroscopic level in an atomic beam. According to a simple resonator model, the tuning instabilities can be ascribed to thermal drifts in the dye-laser system. Problems encountered in attempting a single mode scanning are also described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Applied Physics B Photophysics and Laser Chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.