Abstract
Method 8080, as published in the Second Edition of "Test Methods for Evaluating Solid Waste," EPA Manual SW-846, has been evaluated in a single-laboratory study. The Florisil procedure recommended in Method 8080 for sample cleanup does not separate organochlorine pesticides (OCPs) from the polychlorinated biphenyls (PCBs). Consequently, gas chromatographic analysis of OCPs on the packed column recommended in the method may result in false identifications or in no identifications at all when PCBs are present. Therefore, silica gel was substituted for Florisil, a capillary column was substituted for the packed column, and a sulfur cleanup procedure was incorporated in the method. The revised method was evaluated with samples spiked at 3 concentrations. Precision and accuracy indicate that the revised method can be reliably applied to the determination of OCPs and PCBs in liquid and solid matrixes. Detection limits for liquid matrixes range from 0.02 to 0.09 micrograms/L for OCPs and from 0.5 to 0.9 micrograms/L for PCBs. Detection limits for solid matrixes range from 1 to 6 micrograms/kg for OCPs and from 60 to 70 micrograms/kg for PCBs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.