BackgroundIn this dosimetric study, a dedicated planning tool for single isocenter stereotactic radiosurgery for multiple brain metastases using dynamic conformal arc therapy (DCAT) was compared to standard volumetric modulated arc therapy (VMAT).MethodsTwenty patients with a total of 66 lesions who were treated with the DCAT tool were included in this study. Single fraction doses of 15–20 Gy were prescribed to each lesion. Patients were re-planned using non-coplanar VMAT. Number of monitor units as well as V4Gy, V5Gy and V8Gy were extracted for every plan. Using a density-based clustering algorithm, V10Gy and V12Gy and the volume receiving half of the prescribed dose were extracted for every lesion. Gradient indices and conformity indices were calculated. The correlation of the target sphericity, a measure of how closely the shape of the target PTV resembles a sphere, to the difference in V10Gy and V12Gy between the two techniques was assessed using Spearman’s correlation coefficient.ResultsThe automated DCAT planning tool performed significantly better in terms of all investigated metrics (p < 0.05), in particular healthy brain sparing (V10Gy: median 3.2 cm3 vs. 4.9 cm3), gradient indices (median 5.99 vs. 7.17) and number of monitor units (median 4569 vs. 5840 MU). Differences in conformity indices were minimal (median 0.75 vs. 0.73) but still significant (p < 0.05). A moderate correlation between PTV sphericity and the difference of V10Gy and V12Gy between the two techniques was found (Spearman’s rho = 0.27 and 0.30 for V10Gy and V12Gy, respectively, p < 0.05).ConclusionsThe dedicated DCAT planning tool performed better than VMAT in terms of healthy brain sparing and treatment efficiency, in particular for nearly spherical lesions. In contrast, VMAT can be superior in cases with irregularly shaped lesions.

Highlights

  • In this dosimetric study, a dedicated planning tool for single isocenter stereotactic radiosurgery for multiple brain metastases using dynamic conformal arc therapy (DCAT) was compared to standard volumetric modulated arc therapy (VMAT)

  • For the data evaluated on the plan-wise level (N = 20), the V4Gy, V5Gy and V8Gy of the brain as well as cumulative monitor unit (MU), were superior for all investigated metrics in the Multiple Brain Mets stereotactic radiosurgery (SRS)® (MBSRS) plans (p < 0.05)

  • MBSRS was superior to VMAT for both V10Gy and V12Gy (p < 0.05)

Read more

Summary

Introduction

A dedicated planning tool for single isocenter stereotactic radiosurgery for multiple brain metastases using dynamic conformal arc therapy (DCAT) was compared to standard volumetric modulated arc therapy (VMAT). Whole brain radiotherapy (WBRT) has been the only option for patients with multiple cerebral metastases, whereas patients with a limited number and size of the metastases have been treated with stereotactic radiosurgery (SRS). Guidelines suggested a stereotactic approach for up to 4 metastases [1, 2]. In 2014, Yamamoto et al published a 2014 multicentric prospective trial of 1194 patients treated with GammaKnife SRS. The study compared outcome in patients undergoing multiple SRS treatments for 5–10 brain metastases, as compared to patients with 2–4 metastases.

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.