Abstract

In single input state polarization-sensitive optical coherence tomography (PS-OCT) with high resolution, the imperfections of quarter-wave plate (QWP) and the sensitivity roll-off mismatch between the two detection channels cause unpredictable polarization distortion. We present a correction method based on the Jones matrix modeling of the system. In a single input PS-OCT system working at 840 nm with an axial resolution of ~2.3 μm, the method yielded better estimation of retardation and optic axis orientation with significantly reduced noise level, especially in weakly birefringent samples. Numerical simulations and quantitative imaging of a sample of known birefringence were performed to validate the performance. We further demonstrate the advantages of our approach with birefringence imaging of swine retina, rat aortic wall, and rat esophageal mucosa for potential clinical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.