Abstract

We have prepared fiber-optic pH sensors consisting of a small drop of aminated polystyrene on the tip of a single optical fiber with a core diameter of 100 microns. The sensor is prepared by dipcoating a partially polymerized solution and then completing the polymerization on the fiber. This is followed by amination with diethanolamine. The polymer formulation includes xylene/dodecane to introduce porosity and Kraton G1652, a styrene-ethylene, butylene-styrene, triblock copolymer as a toughening agent. The polymer swells as the amine groups are protonated. This is accompanied by an increase in clarity of the polymer and a decrease in the intensity of light reflected back into the optical fiber. Intensity decreases by over a factor of 2 as the pH is decreased from 8.0 to 6.5. The resulting sensor is small and mechanically stable with a response time of several minutes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.