Abstract

Liquid biopsy is a rapidly emerging field that involves the minimal/non-invasive assessment of signature somatic mutations through the analysis of circulating tumor DNA (ctDNA) shed by tumor cells in bodily fluids. Broadly speaking, the unmet need in liquid biopsy lung cancer detection is the lack of a multiplex platform that can detect a mutation panel of lung cancer genes using a minimum amount of sample, especially for ultra-short ctDNA (usctDNA). Here, we developed a non-PCR and non-NGS-based single-droplet-based multiplexing microsensor technology, "Electric-Field-Induced Released and Measurement (EFIRM) Liquid Biopsy" (m-eLB), for lung cancer-associated usctDNA. The m-eLB provides a multiplexable assessment of usctDNA within a single droplet of biofluid in only one well of micro-electrodes, as each electrode is coated with different probes for the ctDNA. This m-eLB prototype demonstrates accuracy for three tyrosine-kinase-inhibitor-related EGFR target sequences in synthetic nucleotides. The accuracy of the multiplexing assay has an area under the curve (AUC) of 0.98 for L858R, 0.94 for Ex19 deletion, and 0.93 for T790M. In combination, the 3 EGFR assay has an AUC of 0.97 for the multiplexing assay.