Abstract
We report here on first intensity-modulated radiation therapy (IMRT) clinical tests performed at the Gustave Roussy Institute in France using one small (0.117 mm3) synthetic single crystal diamond detector (SCDD). We report the comparison between the point doses evaluated with our detector, with a PTW semiflex air ionization chamber (0.125 cm3) and calculated with the treatment planning system (TPS), respectively. The obtained results show a maximum difference of 2.3% for the diamond detector and of 4.6% for the ionization chamber, as compared with the TPS calculations. These very promising results show the potentiality of chemical vapor deposited SCDD for dosimetry of IMRT fields and opens up the field for diamond dosimeters toward novel applications such as very small beam monitoring.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.