Abstract

We present a method for measuring the magnetic field that allows hyperfine and Zeeman optical pumping, excitation and detection of magnetic resonance by means of a single laser beam with time-modulated ellipticity. This improvement allows us to significantly simplify the Bell-Bloom magnetometric scheme, while retaining its sensitivity. The method does not require the use of radio frequency fields, which is essential when creating arrays of sensors. The results of experimental studies demonstrate the efficiency of the proposed method and its potential applicability in most challenging magnetoencephalographic tasks.