Abstract
Computerized dosimetric studies of single- and double-plane iridium-192 (Ir-192) planar implants were performed. With respect to dose homogeneity, we found that the optimal source and ribbon separation for single-plane implants was 1.0 cm. For double-plane implants, the preferred ribbon and plane separation was 1.5 cm, maintaining a 1-cm separation for the sources. Using these separations, standard dose rate curves for single- and double-plane Ir-192 implants were generated by computer calculations. These standard curves are useful for quickly and fairly accurately estimating the dose from any size planar implant, without requiring more time-consuming individual computer dosimetry. We believe that the curves will prove to be of practical clinical value to physicists and radiotherapists.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.