Abstract

Exposures of an anthropomorphic thorax phantom in posteroanterior projection were made using a focus-film distance of 200 cm and 125 kV with film-screen combinations in class 200 (Curix RP 1L/MR 200) and class 400 (Curix ST-L/Regular). Pathomorphological structures of the lung parenchyma of fine nodular, linear and reticular types were simulated by suitable materials. A ROC analysis including a t-test showed no significant difference in the evaluation of these structural details (p > 0.05). In order to describe image quality for the two film-screen systems, density curves, sigma D-curves and modulation transfer function (MTF) were determined. The only demonstrable difference was reduced MTF of 10-15% for the higher intensification film-screen combination with a resulting poorer resolution of 3.7 l/mm (for the 400 system) compared with 4.3 l/mm for the 200 system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.