Abstract

A tomographic setup that provides the co-registration of photoacoustic (PA) and ultrasound (US) images is presented. For pulse-echo US-tomography laser-induced broadband plane ultrasonic waves are produced by illuminating an optically absorbing target with a short near-infrared laser pulse. Part of the same pulse is frequency doubled and used for the generation of PA waves within the object of interest. The laser-generated plane waves are scattered at the imaging object and measured with the same interferometric detector that also acquires the photoacoustic signals. After collection and separation of the data image reconstruction is done using back-projection resulting in three-dimensional, co-registered PA and US images. The setup is characterized and the resolution in PA and US mode is estimated to be about 85 µm and 40 µm, respectively. Besides measurements on phantoms the performance is also tested on a biological sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.