Abstract

BackgroundQuantification of the longitudinal- and transverse relaxation time in the myocardium has shown to provide important information in cardiac diagnostics. Methods for cardiac relaxation time mapping generally demand a long breath hold to measure either T1 or T2 in a single 2D slice. In this paper we present and evaluate a novel method for 3D interleaved T1 and T2 mapping of the whole left ventricular myocardium within a single breath hold of 15 heartbeats.MethodsThe 3D-QALAS (3D-quantification using an interleaved Look-Locker acquisition sequence with T2 preparation pulse) is based on a 3D spoiled Turbo Field Echo sequence using inversion recovery with interleaved T2 preparation. Quantification of both T1 and T2 in a volume of 13 slices with a resolution of 2.0x2.0x6.0 mm is obtained from five measurements by using simulations of the longitudinal magnetizations Mz. This acquisition scheme is repeated three times to sample k-space. The method was evaluated both in-vitro (validated against Inversion Recovery and Multi Echo) and in-vivo (validated against MOLLI and Dual Echo).ResultsIn-vitro, a strong relation was found between 3D-QALAS and Inversion Recovery (R = 0.998; N = 10; p < 0.01) and between 3D-QALAS and Multi Echo (R = 0.996; N = 10; p < 0.01). The 3D-QALAS method showed no dependence on e.g. heart rate in the interval of 40–120 bpm. In healthy myocardium, the mean T1 value was 1083 ± 43 ms (mean ± SD) for 3D-QALAS and 1089 ± 54 ms for MOLLI, while the mean T2 value was 50.4 ± 3.6 ms 3D-QALAS and 50.3 ± 3.5 ms for Dual Echo. No significant difference in in-vivo relaxation times was found between 3D-QALAS and MOLLI (N = 10; p = 0.65) respectively 3D-QALAS and Dual Echo (N = 10; p = 0.925) for the ten healthy volunteers.ConclusionsThe 3D-QALAS method has demonstrated good accuracy and intra-scan variability both in-vitro and in-vivo. It allows rapid acquisition and provides quantitative information of both T1 and T2 relaxation times in the same scan with full coverage of the left ventricle, enabling clinical application in a broader spectrum of cardiac disorders.

Highlights

  • Quantification of the longitudinal- and transverse relaxation time in the myocardium has shown to provide important information in cardiac diagnostics

  • In this paper we present and evaluate a novel method for 3D interleaved T1 and T2 quantification of the whole left ventricular myocardium within a single breath hold of 15 heartbeats

  • Phantom studies The relation between 3D-QALAS and Inversion Recovery for T1 respectively Multi Echo for T2 was investigated using Pearson correlation coefficient

Read more

Summary

Introduction

Quantification of the longitudinal- and transverse relaxation time in the myocardium has shown to provide important information in cardiac diagnostics. It has become feasible to obtain a quantitative measurement of the signal intensity corresponding to the absolute value of the longitudinal relaxation (T1), transverse relaxation (T2) and proton density in many organs [1] This enables numerical comparison between different tissue compositions and the objective detection of pathological processes by measuring the deviation from the range of normal values. Another important advantage of these quantitative images is the application of synthetic CMR, which uses the absolute parameters T1 and T2 to synthesize any T1-weighted or T2-weighted contrast image [2]. In this way conventional images can be interpreted with simultaneous access to the quantitative parameters from the same acquisition

Methods
Results
Discussion
Conclusion