We demonstrate the use of tone-burst excitation and time-gated spectral analysis for photoacoustic Doppler mapping of flow in an unperturbed vessel phantom and in a vessel with a spatially varying lumen. The method, which mimics pulsed Doppler ultrasound, enables simultaneous measurement of axial position and flow as well as complete characterization of the Doppler spectrum over a wide range of mean velocities (3.5 to 200 mm∕s). To generate the required optical excitation, a continuous cw laser source followed by an external electro-optic modulator is used. Stenoses at various levels are emulated in a C-flex tube with a flowing suspension of micrometer-scale carbon particles. Two-dimensional maps of spectral content versus axial position at different points along the vessel and for various levels of perturbations demonstrate the potential use of the method for characterization of flow irregularities.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.