Abstract

In about a 3 min period, we have simultaneously separated plasma from human whole blood and metered and diluted the plasma using a microchip with an interchannel microstructure. The plasma separation was based on both cross-flow filtration and sedimentation of red blood cells in the microchannels. Metering and diluting operations of the plasma were based on volume control of liquid in the microchannels by syringe pumps. On this microchip, we produced plasma diluted by a factor of 6 from whole blood containing theophylline and we observed very little hemolysis. It is possible to separate plasma from one or just several drops of whole blood by using this microchip.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.