Abstract

A system has been tested that allows simultaneous recording of the retinal response (electroretinogram [ERG]) and the occipital response (visual evoked response [VER]) with focal photic stimulation of the retina under direct observation of the fundus. A helium-neon gas laser is used as a stimulus source. The laser is chopped either by a pen motor or a rotating disc. The laser is attached to a biomicroscope through which the examiner can observe the fundus of the subject during the entire recording session. The optically clear contact lens is made with a flat surface that neutralizes refraction due to the cornea, thereby allowing fundus observation by microscope. Two metal wires mounted inside and outside of the lens serve as the electrode for the ERG. Graticules consisting of concentric circles and radial lines are projected onto the subject's fundus, providing a pattern that the examiner can use to determine the exact location to be stimulated in the fundus. With proper adjustment of stimulus and background illumination, local ERG and VER can be recorded simultaneously by stimulating the macula.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.