Abstract
A procedure for the simultaneous quantitative determination and selective identification of potassium canrenoate and butizide is proposed. The amountof canrenone, the degradation product of potassium canrenoate, is also determined. Elution of these compounds is investigated on LiChrosorb RP-18 and RP-8 columns. The analysis time was shorter on a RP-8 column while the specific identification was still possible. Therefore, the LiChrosorb RP-8 column was chosen. The eluent consisted of an acetonitrile—0.005 M phosphate buffer, pH 4 (45:55) mixture. Quantitative determination was performed at the adsorption maximum of each compound: 286 nm for potassium canrenoate and for canrenone, 271 nm for butizide. This change in wavelength is performed automatically by the computer that controls the spectrophotometric diode array detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.