Abstract

An LC-MS/MS method is presented for the simultaneous quantification of two structurally closely related protein biomarker isoforms, the 22-kDa isoforms of human growth hormone 1 and human growth hormone 2, in human plasma. It is based on multiplexed immunocapture using two monoclonal antibodies immobilized on magnetic beads, tryptic digestion and quantification of two specific signature peptides plus an additional peptide for estimation of total growth hormone related concentrations. A full validation according to international guidelines was performed across the clinically relevant concentration ranges of 0.5 to 50 ng/mL for growth hormone 1, and 2 to 50 ng/mL for growth hormone 2 and demonstrated satisfactory method performance in terms of accuracy, precision, stability and absence of interference. The method’s applicability for routine analysis and its ability to effectively distinguish between GH1 and GH2 was demonstrated by the analysis of plasma samples from pregnant individuals to study the changes in growth hormone levels during pregnancy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.