Abstract

The aim was to develop a fluorescent multiplex array for simultaneously measuring regulated food allergens using specific allergen protein molecules from peanut, tree nut, cow’s milk, egg, soy, fish, shellfish, sesame, mustard and celery. Microspheres coupled to specific monoclonal antibodies were used for allergen detection, with purified allergens as reference standards.Standard curves for 17 allergens covered a 5-log dynamic range. Intra- and inter-assay acceptance criteria were within 70–130% recovery and a CV of ≤15%. Food reference materials contained high levels of their respective major allergens (2000–175,000 µg/g), Similar high allergen levels were found in 10 selected foods analysed using a 9-plex array. Egg, milk, peanut, hazelnut and walnut allergens were detectable in chocolate bars with incurred allergens at 3, 10, 30, and 100 ppm. The multiplex array is an efficient tool for measuring specific food allergens, with applications for risk assessment and standardization of therapeutic products for food allergy.