Abstract

A simple and sensitive liquid chromatography–tandem mass spectrometry (LC–MS/MS) method employing electronspray ionization was developed and validated for quantification of losartan and its carboxylic acid metabolite in human plasma using irbesartan as internal standard (IS). Following a simple pretreatment procedure, the analytes were separated using a gradient mobile phase on reverse phase C18 column. Selected reaction monitoring was specific for losartan, losartan acid and irbesartan. The method validation demonstrated the specificity, lower limit of quantification, accuracy and precision of measurements. The assay exhibited a linear dynamic range of 2.0–400ng/mL for losartan and 1.85–370ng/mL for losartan acid. A run time of 3.5min for each sample made it possible to analyze more than 200 samples per day. The validated method has been successfully used to analyze human plasma samples for application in bioavailability/bioequivalence studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.