Abstract

A cloud point extraction (CPE) method has been developed for the pre-concentration and simultaneous determination of lead [Pb(II)] and tin [Sn(II)] using Acridine Orange as complexing reagent and mediated by nonionic surfactant (Triton X-114) by flame atomic absorption spectrometry (FAAS). The main factors affecting analytical performance of CPE have been investigated in detail. The extracted surfactant-rich phase was diluted with (1.0 mol L−1) nitric acid in methanol, prior to subjecting FAAS. The calibration graphs obtained from Pb(II) and Sn(II) were linear in the concentration ranges of 5–1,000 and 10–5,000 μg L−1 with detection limits of 1.40 and 2.86 μg L−1, respectively. The relative standard deviations for 10 replicates containing 25 μg L−1 of Pb(II) and Sn(II) were 2.15 and 2.50 %, respectively. The analytical procedure was verified by the analysis of the standard reference materials NWTMDA-61.2 (water-trace elements) and NIST SRM 1548a (typical diet). The developed method has been applied to the simultaneous determination of total Pb and Sn in canned food samples including juices, tomato paste, corn, and green pea.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.