Abstract
This study aims to evaluate the utility of simultaneous multislice (SMS) acceleration for routine magnetic resonance neurography (MRN) at 3 T. Patients with multiple sclerosis underwent MRN of the sciatic nerve consisting of a standard fat-saturated T2-weighted turbo spin echo (TSE) sequence using integrated parallel acquisition technique (PAT2) acceleration and 2 T2 TSE sequences using a combination of PAT-SMS acceleration (1) to reduce scan time (PAT2-SMS2; SMS-TSE FAST ) and (2) for time neutral increase of in-plane resolution (PAT1-SMS2; SMS-TSE HR ). Acquisition times were 5:29 minutes for the standard T2 TSE, 3:12 minutes for the SMS-TSE FAST , and 5:24 minutes for the SMS-TSE HR . Six qualitative imaging parameters were analyzed by 2 blinded readers using a 5-point Likert scale and T2 nerve lesions were quantified, respectively. Qualitative and quantitative image parameters were compared, and both interrater and intrarater reproducibility were statistically assessed. In addition, signal-to-noise ratio/contrast-to-noise ratio (CNR) was obtained in healthy controls using the exact same imaging protocol. A total of 15 patients with MS (mean age ± standard deviation, 38.1 ± 11 years) and 10 healthy controls (mean age, 29.1 ± 7 years) were enrolled in this study. CNR analysis was highly reliable (intraclass correlation coefficient, 0.755-0.948) and revealed a significant CNR decrease for the sciatic nerve for both SMS protocols compared with standard T2 TSE (SMS-TSE FAST /SMS-TSE HR , -39%/-55%; P ≤ 0.01). Intrarater and interrater reliability of qualitative image review was good to excellent (κ: 0.672-0.971/0.617-0.883). Compared with the standard T2 TSE sequence, both SMS methods were shown to be superior in reducing pulsatile flow artifacts ( P < 0.01). Ratings for muscle border sharpness, detailed muscle structures, nerve border sharpness, and nerve fascicular structure did not differ significantly between the standard T2 TSE and the SMS-TSE FAST ( P > 0.05) and were significantly better for the SMS-TSE HR than for standard T2 TSE ( P < 0.001). Muscle signal homogeneity was mildly inferior for both SMS-TSE FAST ( P > 0.05) and SMS-TSE HR ( P < 0.001). A significantly higher number of T2 nerve lesions were detected by SMS-TSE HR ( P ≤ 0.01) compared with the standard T2 TSE and SMS-TSE FAST , whereas no significant difference was observed between the standard T2 TSE and SMS-TSE FAST . Implementation of SMS offers either to substantially reduce acquisition time by over 40% without significantly impeding image quality compared with the standard T2 TSE or to increase in-plane resolution for a high-resolution approach and improved depiction of T2 nerve lesions while keeping acquisition times constant. This addresses the specific needs of MRN by providing different imaging approaches for 2D clinical MRN.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.